Biological safety depends on proper cleanup and removal of potentially harmful agents. Disinfection and sterilization are two ways to help ensure biological safety in the laboratory.
The following sections discuss guidelines and procedures for biological disinfection and sterilization:
General Guidelines
Choosing the best method for disinfection and sterilization is very important. The proper method depends on the following:
Once you have chosen the proper method for disinfection or sterilization, follow these guidelines to ensure laboratory safety:
Use the following table to aid in the selection of disinfectants:
| Disinfectant | Uses |
|---|---|
| Alcohols | Ethyl or isopropyl alcohol at 70-80% concentration is a good general purpose disinfectant; not effective against bacterial spores. |
| Phenols | Effective against vegetative bacteria, fungi, and viruses containing lipids; unpleasant odor. |
| Formaldehyde | Concentration of 5-8% formalin is a good disinfectant against vegetative bacteria, spores, and viruses; known carcinogen; irritating odor. |
| Quaternary Ammonium Compounds | Cationic detergents are strongly surface active; extremely effective against lipoviruses; ineffective against bacterial spores; may be neutralized by anionic detergents (i.e., soaps). |
| Chlorine | Low concentrations (50-500 ppm) are active against vegetative bacteria and most viruses; higher concentrations (2,500 ppm) are required for bacterial spores; corrosive to metal surfaces; must be prepared fresh; laundry bleach (5.25% chlorine) may be diluted and used as a disinfectant. |
| Iodine | Recommended for general use; effective against vegetative bacteria and viruses; less effective against bacterial spores; Wescodyne diluted 1 to 10 is a popular disinfectant for washing hands. |
NOTE:
See the Radiation Safety chapter for information pertaining to the use of ultraviolet lights as a method of disinfection.
There are three common methods for sterilizing laboratory materials: wet heat, dry heat, and ethylene oxide gas.
Wet Heat
When used properly, the damp steam heat from an autoclave effectively sterilizes biohazardous waste. Sterilization occurs when contaminated materials reach 15 psi pressure at 250°F or 121°C for at least 30 minutes.
¡IMPORTANT!
For the autoclave process to be effective, sufficient temperature, time, and direct steam contact are essential.
Every Texas State University - San Marcos department that autoclaves biohazardous waste should have written documentation to ensure the waste is sterile. Parameters for sterilization and standard operating procedures should include requirements for verifying sterilization.
Potential problems with wet heat sterilization and autoclaves include the following:
To ensure that all materials are sterile, always test autoclave loads. Remember, however, that some sterilization indicators are incomplete. Autoclave tape, for example, verifies sufficient external temperature exposure, but it does not indicate internal equipment temperature, exposure time, or steam penetration. Thermocouples or other instrumentation can also indicate temperature, but they do not verify sterility. A biological indicator is the most effective monitor to ensure sterility. Commercially available strips or vials of Bacillus species endospores, for example, are suitable biological indicators.
Dry Heat
Dry heat is less effective than wet heat for sterilizing biohazardous materials. Dry heat requires more time (two to four hours) and a higher temperature (320–338°F or 160–170°C) to achieve sterilization. A Bacillus species biological indicator can verify dry heat sterilization.
Ethylene Oxide Gas
Ethylene oxide gas is lethal to all microorganisms. Because it is also a known carcinogen and potentially explosive (freon and carbon dioxide mixtures are stable), minimize your exposure and use extreme care when working with this gas. Ethylene oxide sterilizers and aerators must be properly vented. Ethylene oxide gas is most effective with heat-resistant organisms and heat sensitive equipment. The effectiveness of ethylene oxide gas may be affected by the following: